Notes on the Design of the Liquid Limit Device
Notes on the Design of the Liquid Limit Device
复制标题
DOI:
10.1680/geot.1958.8.2.84
复制
发表时间:
1958-06
期刊:
影响因子:
5.8
通讯作者:
A. Casagrande
中科院分区:
文献类型:
--
作者:
A. Casagrande
Synopsis From experience during the 30 years since he built the first liquid limit device, and from his observations on the use and abuse of this device in many laboratories, the Author concludes that accurate duplication of results requires improved standardization of several details in its design. The design shown in Figs 1 and 2 is recommended as a standard. It maintains the basic dimensions and hardness of base which have been accepted by tacit agreement and have been in use in many countries for about a quarter of a century. In discussing the complications arising from the use of a substantially different hardness of base in the British Standard, the Author suggests as a possible solution replacement of the liquid limit test by a more rational test based on the measurement of shear strength. D'apres l'experience acquise au tours des 30 ans depuis qu'il construisit le premier appareil a limite de liquidite, et d'apres ses observations sur l'emploi et l'abus d'emploi de cet appareil dans de nombreux la...