Limitation of 25-gauge vitrectomy instrumentation in highly myopic eyes
Limitation of 25-gauge vitrectomy instrumentation in highly myopic eyes
复制标题
DOI:
10.3928/15428877-20070901-17
复制
发表时间:
2007-09-01
影响因子:
--
通讯作者:
Stewart, Jay M.
中科院分区:
文献类型:
--
作者:
Singh, Ajay;Fawzi, Arnani A.;Stewart, Jay M.
In highly myopic eyes, a 25-gauge soft-tip needle is unable to remove fluid completely during air-fluid exchange because the needle is too short to reach the back of the elongated globe. In the simulated example, we found that the standard soft-tip cannulated needle is unsuitable for use in eyes with axial lengths greater than 31 mm. Other 25-gauge instruments are similarly unable to reach the back of a long eye.