Meta-analysis of BACTEC MGIT 960 and BACTEC 460 TB, with or without solid media, for detection of mycobacteria
Meta-analysis of BACTEC MGIT 960 and BACTEC 460 TB, with or without solid media, for detection of mycobacteria
复制标题
DOI:
10.1128/jcm.42.5.2321-2325.2004
复制
发表时间:
2004-05-01
影响因子:
9.4
通讯作者:
Mengoli, C
中科院分区:
文献类型:
--
作者:
Cruciani, M;Scarparo, C;Mengoli, C
In a meta-analysis of 10 studies, the BACTEC 960/MGIT and BACTEC 460 systems showed a sensitivity and specificity in detecting mycobacteria (1,381 strains from 14,745 clinical specimens) of 81.5 and 99.6% and 85.8 and 99.9%, respectively. Combined with solid media, the sensitivity of the two systems increased to 87.7 and 89.7%, respectively.