Slit camera focal spot measurement errors in mammography.
Slit camera focal spot measurement errors in mammography.
复制标题
乳房 X 线摄影中的狭缝相机焦点测量误差。
DOI:
--
复制
发表时间:
1995
期刊:
影响因子:
--
通讯作者:
R. L. Tanner
中科院分区:
文献类型:
--
作者:
S. Tang;G. Barnes;R. L. Tanner
Mammography x-ray tube focal spot sizes are routinely measured during acceptance testing and annual performance audits. The National Electrical Manufactures Association (NEMA) recommends the slit camera for this purpose. Investigated were the effect of slit rotational misalignment, tilt misalignment, image film density, film and screen-film image receptors, microscope magnification and reticule accuracy, and observer variation on slit camera focal spot measurements. Our results indicate that small rotational misalignment (< 5 degrees) and tilt misalignment (< 3 degrees) introduce insignificant error. Measured focal spot size increased slightly with image optical density, indicating that for consistent results the image optical density variations should be minimized. Also desirable for accurate field measurements is a high power microscope (25-50x) and a reticule with divisions of < or = 0.02 mm. Screen-film imaging consistently resulted in a slightly smaller measured focal spot size than direct film. The greatest source of error was due to observer variation. Of interest is that reader variability showed a consistent pattern and variation between two measurements by the same observer was much smaller than between observer variation, suggesting that standardized criteria should be established and a method of reader training developed. The length of the focal spot is defined at a reference axis angle specified by the mammography unit manufacturer. Presented is a tabulation of the focal spot geometry and reference axis angles for the majority of mammography units currently and recently marketed in North America.