Gas-chromatographic analysis for cocaine in human plasma, with use of a nitrogen detector.
Gas-chromatographic analysis for cocaine in human plasma, with use of a nitrogen detector.
复制标题
使用氮气检测器对人血浆中的可卡因进行气相色谱分析。
作者:
P. Jatlow;D. Bailey
We describe a procedure for measuring the concentration of cocaine in plasma by gas chromatography, with use of a nitrogen detector. The procedure was tested on patients receiving intranasal cocaine for topical anesthesia. The propyl ester of benzoylecgonine is used as an internal standard. The lower limit of sensitivity of the procedure is 5--10 mug of cocaine per liter of plasma. The coefficients of variation are 5.3% and 3.0% for concentrations of 50 and 100 mug/liter, respectively.