AN AMPEROMETRIC ENZYME ELECTRODE FOR MONITORING BRAIN GLUCOSE IN THE FREELY MOVING RAT
AN AMPEROMETRIC ENZYME ELECTRODE FOR MONITORING BRAIN GLUCOSE IN THE FREELY MOVING RAT
复制标题
DOI:
10.1016/0304-3940(86)90527-6
复制
发表时间:
1986-12-23
影响因子:
2.5
通讯作者:
BARTLETT, PN
中科院分区:
文献类型:
--
作者:
BOUTELLE, MG;STANFORD, C;BARTLETT, PN
Brain glucose concentration was measured with an amperometric enzyme electrode using glucose oxidase (EC 1.1.3.4) irreversibly adsorbed onto an organic conducting salt. The responses of the electrode and its stability both in vitro and in vivo are described. Parallel changes in brain glucose and blood glucose (measured in samples from an implanted intra-atrial cannula) following injections of insulin are reported.