DQE analysis on a dual detector phase x-ray imaging system.

DQE analysis on a dual detector phase x-ray imaging system.
复制标题

双探测器相位 X 射线成像系统的 DQE 分析。

DOI:
10.1088/0031-9155/53/18/021
复制
发表时间:
2008
影响因子:
3.5
通讯作者:
Wu,Xizeng
Wu,Xizeng
中科院分区:
工程技术2区
文献类型:
--
作者:
Zhang,Da;Liu,Hong;Wu,Xizeng

文献摘要

相似文献

This study presents the characterization results of a newly developed dual detector in-line phase x-ray imaging prototype. Comparison of modulation transfer function (MTF), noise power spectrum (NPS) and detective quantum efficiency (DQE) for both detectors was conducted when they worked in the dual detection mode, in which two images are acquired simultaneously at a single exposure. The MTFs of the two detectors are almost identical, showing that the blurring caused by detector1 does not significantly weaken the resolving power of detector2. With a 40 kVp and 4 cm thick BR-12 phantom filtered x-ray beam, the transmittance of detector1 was measured to be 32%. The characteristic response and DQE of the two detectors almost coincide, showing that the two detectors have similar imaging performance under the imaging conditions of this study. The DQE of detector2 at the different source to detector distances (SID) also demonstrate a high level of agreement, implying that the reduced exposure level caused by elongated SID did not degrade the performance significantly. The study validated the design of the dual detection configuration for phase x-ray imaging, which has the potential for improving the accuracy of diagnostics at clinically acceptable radiation doses.