Accurate automatic exposure controller for mammography: design and performance.
Accurate automatic exposure controller for mammography: design and performance.
复制标题
用于乳房 X 线摄影的精确自动曝光控制器:设计和性能。
DOI:
10.1148/radiology.178.2.1987599
复制
发表时间:
1991
期刊:
影响因子:
19.7
通讯作者:
Entine,G
中科院分区:
文献类型:
--
作者:
Frederick,EE;Squillante,MR;Cirignano,LJ;Hahn,RW;Entine,G
An automatic exposure controller has been designed that controls the optical film density for film, screen, and radiographic techniques typically used in mammography to within 0.05 over a range of 1.3-6.7-cm thickness of Lucite. This degree of accuracy is better than that reported for presently available controllers. The detector system consists of four cadmium telluride detectors and involves the use of a control algorithm to read the detectors and turn off the mammography unit at the correct time. This algorithm is implemented by a microprocessor, which also provides the means for a convenient calibration.