Rapid Determination of Vitamin B12 by Inductively Coupled Plasma Mass Spectrometry in Multivitamin Tablets
Rapid Determination of Vitamin B12 by Inductively Coupled Plasma Mass Spectrometry in Multivitamin Tablets
复制标题
DOI:
10.1080/00387010802371387
复制
发表时间:
2008-10
影响因子:
1.7
通讯作者:
Ni Zhang;W. Wang;Kuigang Wang;Zhiying Ding;Zeng Wei
中科院分区:
文献类型:
--
作者:
Ni Zhang;W. Wang;Kuigang Wang;Zhiying Ding;Zeng Wei
ABSTRACT In the current work, a rapid, simple, and sensitive inductively coupled plasma mass spectrometry (ICP-MS) method for determining vitamin B12 in pharmaceutical preparations is proposed. The calibration graph plotted with five concentrations of vitamin B12 was linear with a regression coefficient r2 = 0.9994. The detection limit was 0.05 ng/mL for vitamin B12. At a pump rate of 30 rpm, an analysis cycle of vitamin B12, including sampling and washing, could be accomplished in 0.5 min with the relative standard deviations of 4.2%, 3.5%, and 2.6% for 0.25 ng/mL, 2.5 ng/mL, and 25 ng/mL, respectively. The method has been applied successfully in the determination of vitamin B12 in multivitamin tablets.