A method to stabilize the incident X-ray energy for anomalous diffraction measurements
A method to stabilize the incident X-ray energy for anomalous diffraction measurements
复制标题
一种稳定异常衍射测量的入射 X 射线能量的方法
DOI:
10.1107/s1600577517006312
复制
发表时间:
2017
影响因子:
2.5
通讯作者:
Liu Peng
中科院分区:
文献类型:
--
作者:
Wang Wenjia;Yang Xiaoyun;Chang Guangcai;An Pengfei;Cha Kewen;Dong Yuhui;Liu Peng
A method to calibrate and stabilize the incident X-ray energy for anomalous diffraction data collection is provided and has been successfully used at the single-crystal diffraction beamline 1W2B at the Beijing Synchrotron Radiation Facilities. Employing a feedback loop to control the movement of the double-crystal monochromator, this new method enables the incident X-ray energy to be kept within a 0.2 eV range at the inflection point of the absorption edge.