ULTRASONIC-ATTENUATION AND ABSORPTION IN LIVER-TISSUE
ULTRASONIC-ATTENUATION AND ABSORPTION IN LIVER-TISSUE
复制标题
DOI:
10.1016/0301-5629(83)90089-3
复制
发表时间:
1983-01-01
影响因子:
2.9
通讯作者:
PARKER, KJ
中科院分区:
文献类型:
--
作者:
PARKER, KJ
A large range of values for ultrasonic attenuation and absorption coefficients of tissues are reported in the literature. An important distinction both practically and theoretically is the magnitude of the true absorption, which characterizes the rate of conversion of ultrasonic to thermal energy, as compared with the total attenuation of the ultrasonic signal as it propagates through tissue. The magnitudes of these quantities were studied in bovine liver. Total attenuation was measured, in the range of 1-6 MHz, by both phase sensitive and phase insensitive insertion loss techniques. Ultrasonic absorption was determined by 2 thermal techniques. The standard transient thermoelectric or rate-of-heating method and a new measurement technique based on the temperature decay following a short ultrasonic pulse were employed for the determination of the ultrasonic absorption coefficient. The results demonstrate that the ultrasonic amplitude attenuation and absorption coefficients at low megahertz frequencies are not significantly different in liver. The mean values cluster around 0.05 nepers/cm per MHz (0.4 dB/cm per MHz). The sample-to-sample variation is indicated by the standard deviation in the measurements of 0.01 nepers/cm per MHz (0.09 dB/cm per MHz) or less. In liver tissue, absorption is the dominant feature of attenuation over this frequency range.