Chloride-selective liquid-membrane electrodes based on lipophilic methyl-tri-N-alkyl-ammonium compounds and their applicability to blood serum measurements
Chloride-selective liquid-membrane electrodes based on lipophilic methyl-tri-N-alkyl-ammonium compounds and their applicability to blood serum measurements
复制标题
基于亲脂性甲基三N烷基铵化合物的氯离子选择性液膜电极及其在血清测量中的应用
DOI:
10.1007/bf01201614
复制
发表时间:
1978
影响因子:
5.7
通讯作者:
W. Simon
中科院分区:
文献类型:
--
作者:
K. Hartman;S. Luterotti;H. Osswald;M. Oehme;P. C. Meier;D. Ammann;W. Simon
Several pure methyltrialkylammonium chlorides were synthesized and evaluated as Cl−-selective components in liquid-membrane electrodes. The selectivities, stabilities and lifetimes of these PVC-based electrodes have been shown to be sensitive to the combination of ammonium compound and plasticizer used; the membrane composition has been optimized for Cl−-activity measurements in blood serum. The inherent suitability of this system for applications to blood serum is borne out by the reproducibilities achieved, which compare well with those of popular analytical techniques for Cl−, as long as contact with aqueous calibration solutions is evaded.