Hand-held optical imager (Gen-2): improved instrumentation and target detectability.
Hand-held optical imager (Gen-2): improved instrumentation and target detectability.
复制标题
DOI:
10.1117/1.jbo.17.8.081402
复制
发表时间:
2012-08
影响因子:
3.5
通讯作者:
Jean Gonzalez;J. Decerce;S. Erickson;Sergio Martinez;A. Nunez;M. Roman;Barbara Traub;C. Flores;Seigbeh M. Roberts;Estrella Hernandez;Wenceslao Aguirre;R. Kiszonas;A. Godavarty
中科院分区:
文献类型:
--
作者:
Jean Gonzalez;J. Decerce;S. Erickson;Sergio Martinez;A. Nunez;M. Roman;Barbara Traub;C. Flores;Seigbeh M. Roberts;Estrella Hernandez;Wenceslao Aguirre;R. Kiszonas;A. Godavarty
Hand-held optical imagers are developed by various researchers towards reflectance-based spectroscopic imaging of breast cancer. Recently, a Gen-1 handheld optical imager was developed with capabilities to perform two-dimensional (2-D) spectroscopic as well as three-dimensional (3-D) tomographic imaging studies. However, the imager was bulky with poor surface contact (~30%) along curved tissues, and limited sensitivity to detect targets consistently. Herein, a Gen-2 hand-held optical imager that overcame the above limitations of the Gen-1 imager has been developed and the instrumentation described. The Gen-2 hand-held imager is less bulky, portable, and has improved surface contact (~86%) on curved tissues. Additionally, the forked probe head design is capable of simultaneous bilateral reflectance imaging of both breast tissues, and also transillumination imaging of a single breast tissue. Experimental studies were performed on tissue phantoms to demonstrate the improved sensitivity in detecting targets using the Gen-2 imager. The improved instrumentation of the Gen-2 imager allowed detection of targets independent of their location with respect to the illumination points, unlike in Gen-1 imager. The developed imager has potential for future clinical breast imaging with enhanced sensitivity, via both reflectance and transillumination imaging.