Centering the projection reconstruction trajectory: Reducing gradient delay errors
Centering the projection reconstruction trajectory: Reducing gradient delay errors
复制标题
DOI:
10.1002/mrm.10501
复制
发表时间:
2003-07-01
影响因子:
3.3
通讯作者:
McVeigh, ER
中科院分区:
文献类型:
--
作者:
Peters, DC;Derbyshire, JA;McVeigh, ER
The projection reconstruction (PR) trajectory was investigated for the effect of gradient timing delays between the actual and requested start time of each physical gradient. Radial trajectories constructed with delayed gradients miss the center of k-space in an angularly dependent manner, causing effective echo times to vary with projection angle. The gradient timing delays were measured in phantoms, revealing delays on the x, y, and z gradients which differed by as much as 5 musec. Using this one-time calibration measurement, the trajectories were corrected for gradient delays by addition of compensatory gradient areas to the prephasers of the logical x and y readout gradients. Effective projection-to-projection echo time variability was reduced to less than 1 musec for all imaging orientations. Using corrected trajectories, artifacts were reduced in phantom images and in volunteer studies. This correction should potentiate greater clinical use of the PR trajectory. Magn Reson Med 50:1-6, 2003. Published 2003 Wiley-Liss, lnc.