Performance assessment of an opto-fluidic phantom mimicking porcine liver parenchyma.
Performance assessment of an opto-fluidic phantom mimicking porcine liver parenchyma.
复制标题
模拟猪肝实质的光流控模型的性能评估。
DOI:
10.1117/1.jbo.17.7.077008
复制
发表时间:
2012
影响因子:
3.5
通讯作者:
Coté,GerardL
中科院分区:
文献类型:
--
作者:
Akl,TonyJ;King,TravisJ;Long,Ruiqi;McShane,MichaelJ;NanceEricson,M;Wilson,MarkA;Coté,GerardL
An implantable, optical oxygenation and perfusion sensor to monitor liver transplants during the two-week period following the transplant procedure is currently being developed. In order to minimize the number of animal experiments required for this research, a phantom that mimics the optical, anatomical, and physiologic flow properties of liver parenchyma is being developed as well. In this work, the suitability of this phantom for liver parenchyma perfusion research was evaluated by direct comparison of phantom perfusion data with data collected fromin vivoporcine studies, both using the same prototype perfusion sensor.In vitroperfusion and occlusion experiments were performed on a single-layer and on a three-layer phantom perfused with a dye solution possessing the absorption properties of oxygenated hemoglobin. While both phantoms exhibited response patterns similar to the liver parenchyma, the signal measured from the multilayer phantom was three times higher than the single layer phantom and approximately 21 percent more sensitive toin vitrochanges in perfusion. Although the multilayer phantom replicated thein vivoflow patterns more closely, the data suggests that both phantoms can be usedin vitroto facilitate sensor design.