Muscle compression improves reliability of ultrasound echo intensity.
Muscle compression improves reliability of ultrasound echo intensity.
复制标题
肌肉压缩提高了超声回波强度的可靠性。
DOI:
10.1002/mus.25779
复制
发表时间:
2018
期刊:
影响因子:
3.4
通讯作者:
Anthony,BrianW
中科院分区:
文献类型:
--
作者:
Pigula-Tresansky,AnneJ;Wu,JimS;Kapur,Kush;Darras,BasilT;Rutkove,SewardB;Anthony,BrianW
IntroductionMuscle echo intensity has been shown to correlate with disease status in muscle disorders, including Duchenne muscular dystrophy (DMD). We report the effect of sonographer‐applied load on measurements of muscle echo intensity.MethodsQuadriceps ultrasound scans were performed on 22 healthy boys and 16 boys with DMD between the ages of 2.2 and 15.3 years. Transducer contact force was increased linearly from 1.5 to 10 N, and echo intensity was measured throughout.ResultsEcho intensity increased linearly with strain at a rate of 42 (95% confidence interval [CI]: 21–63) and 74 (95% CI: 49–98) in the healthy and DMD populations, respectively. Echo intensity reliability was moderate at low strain (intraclass correlation coefficient [ICC] = 0.82) and was improved at high strain (ICC = 0.92).DiscussionSonographer‐applied load introduces error in measurements of echo intensity, but it can be minimized by measuring echo intensity at near‐maximal levels of compression.Muscle Nerve57: 423–429, 2018