Modified Look-Locker T1 evaluation using Bloch simulations: Human and phantom validation
Modified Look-Locker T1 evaluation using Bloch simulations: Human and phantom validation
复制标题
DOI:
10.1002/mrm.24251
复制
发表时间:
2013-02-01
影响因子:
3.3
通讯作者:
Bluemke, David A.
中科院分区:
文献类型:
--
作者:
Gai, Neville D.;Stehning, Christian;Bluemke, David A.
Modified Look-Locker imaging is frequently used for T1 mapping of the myocardium. However, the specific effect of various MRI parameters (e.g., encoding scheme, modifications of flip angle, heart rate, T2, and inversion times) on the accuracy of T1 measurement has not been studied through Bloch simulations. In this work, modified Look-Locker imaging was characterized through a numerical solution for Bloch equations. MRI sequence parameters that may affect T1 accuracy were systematically varied in the simulation. For validation, phantoms were constructed with various T2 and T1 times and compared with Bloch equation simulations. Human volunteers were also evaluated with various pulse sequences parameters to assess the validity of the numerical simulations. There was close agreement between simulated T1 times and T1 times measured in phantoms and volunteers. Lower T2 times (i.e.,