Laser nephelometry and radial immunodiffusion compared for immunoglobulin quantification in pathological sera.
Laser nephelometry and radial immunodiffusion compared for immunoglobulin quantification in pathological sera.
复制标题
激光比浊法和径向免疫扩散法对病理血清中免疫球蛋白的定量进行了比较。
DOI:
10.1093/clinchem/28.1.180
复制
发表时间:
1982
影响因子:
9.3
通讯作者:
R. Creyssel
中科院分区:
文献类型:
--
作者:
H. Cloppet;A. Francina;H. Coquelin;Y. Boucaud;P. Hutinel;R. Creyssel
We evaluated nephelometers from Behring, Hyland, and Beckman for IgG, IgA, and IgM quantitation in sera from patients with monoclonal gammopathies. The intra-batch precision of each instrument for each immunoglobulin class and for different concentrations of the same immunoglobulin was compared to the one obtained with the radial immunodiffusion method. No nephelometer showed a clearly better precision. The correlation with cellulose acetate electrophoresis was good for each of the three nephelometers. The mean value by the radial immunodiffusion method was higher than corresponding determinations by nephelometry.