Determination of Ranitidine and Salbutamol by Flow Injection Analysis with Chemiluminescence Detection
Determination of Ranitidine and Salbutamol by Flow Injection Analysis with Chemiluminescence Detection
复制标题
流动注射化学发光检测法测定雷尼替丁和沙丁胺醇
DOI:
10.1016/s0003-2670(98)00751-x
复制
发表时间:
1999
影响因子:
6.2
通讯作者:
S. Lewis
中科院分区:
文献类型:
--
作者:
N. W. Barnett;B. Hindson;S. Lewis
Flow injection analysis methodology is reported for the determination of ranitidine and salbutamol (in pharmaceutical formulations) based on their chemiluminescent reactions with tris(2,2′-bipyridyl)ruthenium(III) and acidic potassium permanganate, respectively. While the log–log calibration function for ranitidine showed significant non linearity, the precision (measured as relative standard deviation) was 1.7% (n=6 1×10−5M) and the detection limit (3σ blank) was 6×10−7M. A linear log–log calibration function was obtained for salbutamol, the precision being 2.7% (n=6; 1×10−7M) with a detection limit (signal to noise ratio=3) of 2.5×10−8M.