INDOCYANINE GREEN FLUORESCENCE ANGIOGRAPHY
INDOCYANINE GREEN FLUORESCENCE ANGIOGRAPHY
复制标题
吲哚青绿荧光血管造影
DOI:
10.1111/j.1755-3768.1980.tb08294.x
复制
发表时间:
1980
影响因子:
3.4
通讯作者:
R. Flower
中科院分区:
文献类型:
--
作者:
L. Hyvärinen;R. Flower
Indocyanine green (ICG) fluorescence angiography has been further refined for use in both laboratory and clinical investigations. In the present modification of the Zeiss fundus camera all lenses except the aspherical objective lens have been specially antireflection coated to increase light transmission in the spectral region around 800 nm. A 300 watt indium iodide lamp continuous light source has replaced the conventional xenon flash lamp. This light source produces a retinal irradiance of 265 mw, and therefore restricts retinal exposure time to 11.9 seconds, but that time is more than adequate to record passage of dye through the choroid. Spatial resolution of the fundus on the film has been increased from 11.7 microns to 7.4 microns.