Intra- and inter-operator repeatability of myocardial blood flow and myocardial flow reserve measurements using rubidium-82 pet and a highly automated analysis program
Intra- and inter-operator repeatability of myocardial blood flow and myocardial flow reserve measurements using rubidium-82 pet and a highly automated analysis program
复制标题
DOI:
10.1007/s12350-010-9225-3
复制
发表时间:
2010-08-01
影响因子:
2.4
通讯作者:
deKemp, Robert A.
中科院分区:
文献类型:
--
作者:
Klein, Ran;Renaud, Jennifer M.;deKemp, Robert A.
Background. Changes in myocardial blood flow between rest and stress states are commonly used to diagnose coronary artery disease. Relative myocardial perfusion imaging (MPI) is used routinely while myocardial blood flow quantification (MBF) may improve the sensitivity for detection of early disease. The ratio of flow at stress and rest (S/R) and their difference (S-R) have both been proposed as a means to detect regions with reduced myocardial flow reserve (MFR). In this study, we describe a highly automated method to calculate regional and global rest, stress, S/R, and S-R polar maps of the left ventricle myocardium.Methods. We measured the inter- and intra-operator variability using two randomized datasets (n = 30 each) for each of two operators (novice and expert) with correlation and Bland-Altman reproducibility coefficient (RPC%) analyses.Results. S-R MBF had less inter- operator dependent variability than S/R (RPC% = 5.0% vs 12.6%, P