Electronic Detection Strategies for a MEMS-Based Biosensor
Electronic Detection Strategies for a MEMS-Based Biosensor
复制标题
DOI:
10.1109/jmems.2012.2219294
复制
发表时间:
2013-04-01
影响因子:
2.7
通讯作者:
McNeil, Calum
中科院分区:
文献类型:
--
作者:
Burnett, Richie;Harris, Alun;McNeil, Calum
This work reports on the development of control electronics for a resonant mode biosensor. The laboratory demonstrated sensor initially showed a signal-to-noise ratio of -100 dB. With the application of half-frequency drive, frequency down conversion, appropriate filtering, and digital signal processing, a cost-effective electronic solution demonstrated a signal-to-noise ratio of +30 dB. This paper highlights important aspects in the design of integrated solutions to microelectromechanical systems-based sensors.