Baseband Feedback Frequency-Division Multiplexing with Low-Power dc-SQUIDs and Digital Electronics for TES X-Ray Microcalorimeters
Baseband Feedback Frequency-Division Multiplexing with Low-Power dc-SQUIDs and Digital Electronics for TES X-Ray Microcalorimeters
复制标题
用于 TES X 射线微量热计的低功耗 dc-SQUID 和数字电子器件的基带反馈频分复用
DOI:
10.1007/s10909-013-1040-1
复制
发表时间:
2014
影响因子:
2
通讯作者:
T. Miyazaki
中科院分区:
文献类型:
--
作者:
K. Sakai;Y. Takei;R. Yamamoto;N. Y. Yamasaki;K. Mitsuda;M. Hidaka;S. Nagasawa;S. Kohjiro;T. Miyazaki
We are developing frequency-division multiplexing (FDM) systems with baseband feedback for TES X-ray microcalorimeter arrays for use with the DIOS mission. To meet the requirement of limited cooling capacity at cryogenic temperatures, we developed low-power and FDM-optmized dc-SQUIDs. To make maximum use of the SQUIDs, we also developed digital electronics using FPGA evaluation boards and ADC/DAC FMC daughter cards, and evaluated signal-to-noise ratios and gain-bandwidth products.
DOI:
--
发表时间:
2006
期刊:
IEICE Transactions on Electronics E89-C
影响因子:
--
作者:
N.Y.Yamasaki;Y.Takei;K.Masui;K.Mitsuda;T.Morooka;S.Nakayama
通讯作者: