Current-based transthoracic defibrillation

Current-based transthoracic defibrillation
复制标题

DOI:
10.1016/s0002-9149(96)90062-4
复制
发表时间:
1996-11-15
影响因子:
2.8
通讯作者:
Charbonnier, F
Charbonnier, F
中科院分区:
医学3区
文献类型:
--
作者:
Kerber, RE;Kieso, RA;Charbonnier, F

文献摘要

被引文献

相似文献

This study examines in a prospective, multicenter trial the feasibility and advantage of current-based, transthoracic defibrillation. Current-based, damped, sinusoidal waveform shocks of 18, 25, 30, 35, or 40 amperes (A) were administered beginning with 25 A for polymorphic ventricular tachycardia (VT) and ventricular fibrillation (VF) or 18 A for monomorphic VT; success rates were compared with those of energy-based shocks beginning at 200 J for VF/polymorphic VT and 100 J for VT. The current-based shocks were delivered from custom-modified defibrillators that determined impedance in advance of any shock using a ''test-pulse'' technique; the capacitor then charged to the exact energy necessary to deliver the operator-selected current against the impedance determined by the defibrillator. Three hundred sixty-two patients received >1 shock for VF, polymorphic VT, or monomorphic VT: 569 current- based shocks and 420 energy-based shocks. Current-based shocks of 35/40 A achieved success rates of up to 74% for VF/polymorphic VT; 30 A shocks terminated 88% of monomorphic VT episodes. Energy-based shocks of 300 J terminated 72% of VF/polymorphic VT; 200-J shocks terminated 89% of monomorphic VT. We could not demonstrate a significant increase in the success rate of current-based shocks over energy-based shocks for patients with high transthoracic impedance; this may be due to inadequate sample size. Thus, current-based defibrillation is clinically feasible and effective. A larger study will be needed to test whether current-based defibrillation is superior to energy-based defibrillation. (C) 1996 by Excerpta Medica, Inc.