使用 Charnley 柄几何形状和改进的 Charnley 柄几何形状开发了具有近端柄-骨水泥粘合的骨水泥股骨髋关节组件的非线性三维有限元模型,该几何形状在柄的远端三分之二上具有圆柱形横截面(远端圆形)。当近端柄-骨水泥界面脱粘时,Charnley 柄和 Distal-Round 柄的近端骨水泥套的峰值拉应力分别增加 63% 和 74%。具有近端粘结的柄-骨水泥界面上的剪切应力,远端圆形柄比 Charnley 柄大 29%。近端柄-骨水泥界面脱粘后,远端圆形柄的骨水泥套中的峰值拉应力比 Charnley 柄大 15%。结果表明,通过使用近端柄-骨水泥粘合,Charnley 柄和远端圆形柄的近端骨水泥套内的应力可显着降低。然而,由于剪切应力增加,远端圆形柄脱粘的风险可能更高,并且一旦脱粘,远端圆形柄由于水泥套失效而进一步松动的风险也将更高。
Nonlinear, three-dimensional, finite element models of cemented femoral hip components with a proximal stem-cement bond were developed with use of a Charnley stem geometry and a modified Charnley stem geometry that had a cylindrical cross section over the distal two-thirds of the stem (Distal-Round). Peak tensile stresses in the proximal cement mantle increased 63 and 74% for the Charnley and Distal-Round stems, respectively, when the proximal stem-cement interface was debonded. The shear stresses over the stem-cement interface with a proximal bond were 29% larger for the Distal-Round stem than for the Charnley stem. After the proximal stem-cement interface was debonded, the peak tensile stresses in the cement mantle were 15% larger for the Distal-Round stem than for the Charnley stem. The results illustrate that stresses within the proximal cement mantle could be substantially reduced for both Charnley and Distal-Round stems through use of a proximal stem-cement bond. However, the risk of debonding may be higher for the Distal-Round stem because of increased shear stresses, and once debonded the risk of further loosening due to failure of the cement mantle would also be higher for the Distal-Round stem.