Beam scanning system for a clinical pi-meson beam
Beam scanning system for a clinical pi-meson beam
复制标题
用于临床 pi 介子束的光束扫描系统
DOI:
10.1063/1.1138940
复制
发表时间:
1986
期刊:
影响因子:
--
通讯作者:
L. Skarsgard
中科院分区:
文献类型:
--
作者:
G. Lam;R. Harrison;R. Kornelsen;L. Skarsgard
A beam delivery system for large uniform dose fields suitable for cancer radiotherapy has been constructed for the TRIUMF pion treatment facility using discrete step beam scanning in three dimensions. The patient on a couch is scanned laterally relative to a pencil beam in a raster fashion. At each spot, the depth dose profile is delivered by range modulations using a range shifter. The operations of the scanning couch and range shifter are synchronized by the use of beam monitor pulses from a beam intensity integrating system. The system is hence versatile enough to handle treatment volume of highly irregular shape and yet is immune to fluctuations in accelerator output.