Linear Accuracy of Cone Beam CT Derived 3D Images
Linear Accuracy of Cone Beam CT Derived 3D Images
复制标题
DOI:
10.2319/122407-599.1
复制
发表时间:
2009-01-01
影响因子:
3.4
通讯作者:
Farman, Allan G.
中科院分区:
文献类型:
--
作者:
Brown, April A.;Scarfe, William C.;Farman, Allan G.
Objective: To compare the in vitro reliability and accuracy of linear measurements between cephalometric landmarks on cone beam computed tomography (CBCT) 3D volumetric images with varying basis projection images to direct measurements on human skulls.Materials and Methods: Sixteen linear dimensions between 24 anatomic sites marked on 19 human skulls were directly measured. The skulls were imaged with CBCT (i-CAT, Imaging Sciences International, Hatfield, Pa) at three settings: (a) 153 projections, (b) 306 projections, and (c) 612 projections. The mean absolute error and modality mean (+/- SD) of linear measurements between landmarks on volumetric renderings were compared to the anatomic truth using repeated measures general linear model (P