Continuous outflow devices for macular translocation with 360-degree retinotomy.
Continuous outflow devices for macular translocation with 360-degree retinotomy.
复制标题
用于 360 度视网膜切开术的黄斑转位的连续流出装置。
DOI:
10.1016/s0002-9394(02)01931-1
复制
发表时间:
2003
影响因子:
4.2
通讯作者:
Stanley Chang
中科院分区:
文献类型:
--
作者:
O. Çekiç;M. Ohji;Y. Tano;Stanley Chang
PURPOSESeparation of the retina from retinal pigment epithelium is the most difficult step in macular translocation surgery. To complete this critical step, we have developed two specially designed continuous outflow needles, a rigid needle for creating retinal detachment and a liquid perfluorocarbon injection needle with a soft tip.DESIGNExperimental study, new instrument development.METHODSTo permit continuous outflow during separation of the retina from the retinal pigment epithelium, the shaft of the continuous outflow needle is smaller than 20 gauge in diameter to allow continuous flow of fluid from the vitreous cavity.RESULTSThese instruments have been used successfully in 54 eyes that underwent macular translocation with 360-degree retinotomy. The rigid needle creates safe retinal detachment by keeping intraocular pressure lower, which is strictly needed for this step. The injection needle facilitates injection of perfluorocarbon liquid while manipulating the retina gently with its soft tip.CONCLUSIONContinuous outflow instruments are useful tools because they facilitate critical steps in the delicate process of macular translocation surgery.