Breast ultrasound elastography-Results of 193 breast lesions in a prospective study with histopathologic correlation

Breast ultrasound elastography-Results of 193 breast lesions in a prospective study with histopathologic correlation
复制标题

DOI:
10.1016/j.ejrad.2009.08.026
复制
发表时间:
2011-03-01
影响因子:
3.3
通讯作者:
Schreer, I.
Schreer, I.
中科院分区:
医学3区
文献类型:
--
作者:
Schaefer, F. K. W.;Heer, I.;Schreer, I.

文献摘要

被引文献

相似文献

目的:评价超声弹性成像对乳腺肿块的诊断价值。材料与方法:采用日本日立公司EUB8500型超声诊断仪和7.5~13 MHz的线阵探头,对乳腺肿块的超声弹性成像进行分析。参考标准为细胞学(细针抽吸)或组织学(核心活检)。弹性评分按6分彩色分级(Ueno分级,1-3=良性,4-5=恶性)。根据BI-RADS分类对常规B型超声(US)所见进行分类。结果:恶性病变的平均弹性评分为4.1+/-0.9,良性病变为2.1+/-1.0(p<0.001)。以弹性评分3~4为最佳分界点,敏感度为96.9%,特异度为76%。以BI-RADS4~5作为常规超声诊断的最佳界点,其敏感性为57.8%,特异性为96.1%。与常规超声相比,弹性成像具有更高的敏感性和更低的特异性,但有2个弹性评分为1的病变为假阴性,而BI-RADS 1~3分的病变均为假阴性。ROC曲线弹性图为0.884,常规超声为0.820(p<0.001)。BI-RADS分级的加权Kappa值为0.784/0.634,弹性评分为0.720/0.561。结论:在我们的研究背景下,弹性成像在乳腺癌的检测中没有取代传统B超的潜力,但可以补充传统的US以提高诊断性能。(C)2009爱思唯尔爱尔兰有限公司。保留所有权利。
Purpose: To evaluate the diagnostic performance of ultrasound elastography in breast masses.Material and methods: 193 lesions (129 benign, 64 malignant) were analyzed with the EUB 8500 Logos-ultrasonic-unit (Hitachi Medical, Japan) and a linear-array-transducer of 7.5-13-MHz. Standard of reference was cytology (FNAfine needle aspiration) or histology (core biopsy). The elastic-score was classified according to a 6-point colour-scale (Ueno classification; 1-3 = benign, 4-5 = malignant). Conventional B-mode ultrasound (US) findings were classified according to the BI-RADS classification. Statistical analysis included sensitivity, specificity, ROC-analysis and kappa-values for intra-/interobserver reliability.Results: The mean score for elasticity was 4.1 +/- 0.9 for malignant lesions, and 2.1 +/- 1.0 for benign lesions (p < 0.001). With a best cut-off point between elasticity scores 3 and 4, sensitivity was 96.9%, and specificity 76%. Setting a best cut-off point for conventional US between BI-RADS 4 and 5, sensitivity was 57.8%, and specificity 96.1%. Elastography provided higher sensitivity and lower specificity than conventional US, but two lesions with elasticity score 1 were false negative, whereas no lesion scored BI-RADS 1-3 were false negative. ROC-curve was 0.884 for elastography, and 0.820 for conventional US (p < 0.001). Weighted kappa-values for intra-/interobserver reliability were 0.784/0.634 for BI-RADS classification, and 0.720/0.561 for elasticity scores.Conclusion: In our study setting, elastography does not have the potential to replace conventional B-mode US for the detection of breast cancer, but may complement conventional US to improve the diagnostic performance. (C) 2009 Elsevier Ireland Ltd. All rights reserved.