Comparison of small field of view gamma camera systems for scintimammography
Comparison of small field of view gamma camera systems for scintimammography
复制标题
DOI:
10.1097/00006231-200505000-00008
复制
发表时间:
2005-05-01
影响因子:
1.5
通讯作者:
Collins, DA
中科院分区:
文献类型:
--
作者:
Hruska, CB;O'Connor, MK;Collins, DA
Objective To compare the performance characteristics of three compact gamma camera systems for scintimammography.Methods A multi-crystal Nal system (Dilon 6800), a multicrystal CsI system (Digirad 2020tc Imager) and a prototype cadmium zinc telluride (CZT) detector were evaluated using a breast phantom with simulated tumours of 1.8-9.8 mm in diameter and a tumour to background ratio of 5.6: 1. A conventional Anger-based gamma camera was also evaluated for comparative purposes.Results The conventional gamma camera had the highest intrinsic sensitivity of the four systems and utilized the highest resolution collimator. However, the prototype CZT and the CsI systems yielded the best spatial resolution over the range 0-5cm from the collimator face and better visualized small (5-7 mm) tumours.Conclusion In addition to the advantage of close proximity to the breast, two of the three compact detectors offered superior performance to a conventional gamma camera for scintimammography. (c) 2005 Lippincott Williams & Wilkins.