Three-dimensional analysis of computed tomography-based navigation system for total knee arthroplasty: the accuracy of computed tomography-based navigation system.
Three-dimensional analysis of computed tomography-based navigation system for total knee arthroplasty: the accuracy of computed tomography-based navigation system.
复制标题
全膝关节置换术计算机断层扫描导航系统的三维分析:计算机断层扫描导航系统的准确性。
DOI:
10.1016/j.arth.2008.07.007
复制
发表时间:
2009
期刊:
影响因子:
--
通讯作者:
Y. Iwamoto
中科院分区:
文献类型:
--
作者:
H. Mizu;S. Matsuda;H. Miura;H. Higaki;K. Okazaki;Y. Iwamoto
We evaluated the postoperative alignment of 37 primary total knee arthroplasties performed using a computed tomography–based navigation system (Vector Vision Knee 1.5; Brain Lab, Germany) with a new 3-dimensional analysis. The mean coronal femoral angle was 89.0° ± 1.4° (85.5°-92.8°), and the coronal tibial component was 89.2° ± 1.0° (87.4°-91.6°). The hip-knee-ankle angle was observed to be 178.2° ± 1.5° (173.9°-181.8°). The external rotational alignment of the femoral component relative to the surgical epicondylar axis was −0.5° ± 1.7° (−3.2° to 3.4°). The results demonstrated that a computed tomography–based navigation system provided a reasonably satisfactory component alignment. The discrepancy between the 2-dimensional and 3-dimensional evaluations was 0.3° ± 1.8° (−2.7° to 3.4°). Three-dimensional analysis is necessary to evaluate the accuracy of the navigation system.