Simultaneous multiple injection in monosegmented flow analysis
Simultaneous multiple injection in monosegmented flow analysis
复制标题
DOI:
10.1016/s0003-2670(98)00321-3
复制
发表时间:
1998-10-05
影响因子:
6.2
通讯作者:
Raimundo, IM
中科院分区:
文献类型:
--
作者:
Brito, VO;Raimundo, IM
A flow approach for simultaneous multiple injection (SMI) in monosegmented flow analysis (MSFA) is described. In this approach, the sample and the reagent (or other solution, such as a diluting fluid) are simultaneously injected into the reaction coil of a monosegmented flow analyser. The monosegment is homogenised while it is carried towards the detector. The sample dilution procedure is not based on a gradient concentration pattern and dilution factors up to 150 were obtained by using hydrodynamic sampling. The system was applied to perform sample dilution in the determination of glucose in blood plasma. The results showed good correlation with those obtained by the Clinical Hospital of UNICAMP. The SMI approach was also applied to add reagent to the sample in nitrite determination in natural water, providing a methodology that has a sampling frequency of 72 h(-1), a relative standard deviation of