Performance of MAP Reconstruction for Hot Lesion Detection in Whole-Body PET/CT: An Evaluation With Human and Numerical Observers
Performance of MAP Reconstruction for Hot Lesion Detection in Whole-Body PET/CT: An Evaluation With Human and Numerical Observers
复制标题
DOI:
10.1109/tmi.2008.927349
复制
发表时间:
2009-01-01
影响因子:
10.6
通讯作者:
Verscuren, Raf
中科院分区:
文献类型:
--
作者:
Nuyts, Johan;Michel, Christian;Verscuren, Raf
For positron emission tomography (PET) imaging, different reconstruction methods can be applied, including maximum likelihood (ML) and maximum a posteriori (MAP) reconstruction. Postsmoothed ML images have approximately position and object independent spatial resolution, which is advantageous for (semi-) quantitative analysis. However, the complex object dependent smoothing obtained with MAP might yield improved noise characteristics, beneficial for lesion detection. In this contribution, MAP and postsmoothed ML are compared for hot spot detection by human observers and fly the channelized Hotelling observer (CHO). The study design was based on the "multiple alternative forced choice" approach. For the MAP reconstruction, the relative difference prior was used. For postsmoothed ML, a Gaussian smoothing kernel was used. Both the human observers and the CHO performed slightly better on MAP images than on postsmoothed ML images. The average CHO performance was similar to the best human performance. The CHO was then applied to evaluate the performance of priors with reduced penalty for large differences. For these priors, a poorer detection performance was obtained.