Bolometric technique for high-resolution broadband microwave spectroscopy of ultra-low-loss samples
Bolometric technique for high-resolution broadband microwave spectroscopy of ultra-low-loss samples
复制标题
用于超低损耗样品高分辨率宽带微波光谱的测辐射热技术
DOI:
10.1063/1.1633001
复制
发表时间:
2003
影响因子:
1.6
通讯作者:
W. Hardy
中科院分区:
文献类型:
--
作者:
P. Turner;D. Broun;S. Kamal;M. Hayden;J. Bobowski;R. Harris;D. C. Morgan;J. Preston;D. Bonn;W. Hardy
A novel low-temperature bolometric method has been devised and implemented for high-precision measurements of the microwave surface resistance of small single-crystal platelet samples having very low absorption, as a continuous function of frequency. The key to the success of this nonresonant method is the in situ use of a normal metal reference sample that calibrates the absolute rf field strength. The sample temperature can be controlled independently of the 1.2 K liquid-helium bath, allowing for measurements of the temperature evolution of the absorption. However, the sensitivity of the instrument decreases at higher temperatures, placing a limit on the useful temperature range. Using this method, the minimum detectable power at 1.3 K is 1.5 pW, corresponding to a surface resistance sensitivity of ≈1 μΩ for a typical 1 mm×1 mm platelet sample.