PEEK-Halo effect in interbody fusion

PEEK-Halo effect in interbody fusion
复制标题

DOI:
10.1016/j.jocn.2015.07.017
复制
发表时间:
2016-02-01
影响因子:
2
通讯作者:
Mobbs, Ralph J.
Mobbs, Ralph J.
中科院分区:
医学4区
文献类型:
--
作者:
Phan, Kevin;Hogan, Jarred A.;Mobbs, Ralph J.

文献摘要

被引文献

相似文献

最近的发展已经看到聚[芳基醚醚酮](PEEK)越来越多地用于椎体融合。改进PEEK的更新颖的方法包括引入钛PEEK(Ti-PEEK)复合材料和涂层。本文旨在描述PEEK基植入物的潜在并发症,该并发症与周围骨结合较差有关,产生了在Ti-PEEK复合植入物中未观察到的“PEEK-Halo”效应。我们展示了两名接受前路腰椎椎间融合术(ALIF)的患者的图像。第1例患者使用PEEK植入物接受L5/S1 ALIF,而第2例患者使用Ti-PEEK复合植入物接受L4/L5 ALIF。使用CT成像寻找骨整合的证据,并使用羊胫骨模型的组织学制备物进行确认。PEEK-光晕效应通过CT成像上PEEK植入物和植骨之间的光晕效应来证明。这种现象继发于PEEK植入物的骨结合不良。在接受Ti-PEEK复合移植物的第二例患者中未证实PEEK-Halo效应。羊模型中PEEK与Ti-PEEK植入物的移植物/骨界面的组织学分析进一步证实PEEK植入物的骨结合较差。总之,PEEK-Halo效应继发于PEEK植入物插入后相邻椎体终板处PEEK的最小骨整合。在Ti-PEEK植入物中未观察到这种效应,这可能支持钛在改善PEEK基质骨-植入物界面方面的作用。(C)2015爱思唯尔有限公司版权所有。
Recent developments have seen poly[aryl-ether-ether-ketone] (PEEK) being increasingly used in vertebral body fusion. More novel approaches to improve PEEK have included the introduction of titanium PEEK (Ti-PEEK) composites and coatings. This paper aims to describe a potential complication of PEEK based implants relating to poorer integration with the surrounding bone, producing a "PEEK-Halo" effect which is not seen in Ti-PEEK composite implants. We present images from two patients undergoing anterior lumbar interbody fusion (ALIF). The first patient underwent an L5/S1 ALIF using a PEEK implant whilst the second patient underwent L4/L5 ALIF using a Ti-PEEK composite implant. Evidence of osseointegration was sought using CT imaging and confirmed using histological preparations of a sheep tibia model. The PEEK-Halo effect is demonstrated by a halo effect between the PEEK implant and the bone graft on CT imaging. This phenomenon is secondary to poor osseointegration of PEEK implants. The PEEK-Halo effect was not demonstrated in the second patient who received a Ti-PEEK composite graft. Histological analysis of graft/bone interface surfaces in PEEK versus Ti-PEEK implants in a sheep model further confirmed poorer osseointegration of the PEEK implant. In conclusion, the PEEK-Halo effect is seen secondary to minimal osseointegration of PEEK at the adjacent vertebral endplate following a PEEK implant insertion. This effect is not seen with Ti-PEEK implants, and may support the role of titanium in improving the bone-implant interface of PEEK substrates. (C) 2015 Elsevier Ltd. All rights reserved.