Energy resolution and high count rate performance of superconducting tunnel junction x-ray spectrometers
Energy resolution and high count rate performance of superconducting tunnel junction x-ray spectrometers
复制标题
超导隧道结X射线光谱仪的能量分辨率和高计数率性能
DOI:
10.1063/1.1148474
复制
发表时间:
1998
影响因子:
1.6
通讯作者:
A. Barfknecht
中科院分区:
文献类型:
--
作者:
M. Frank;L. J. Hiller;J. Grand;C. Mears;S. Labov;M. Lindeman;H. Netel;D. Chow;A. Barfknecht
We present experimental results obtained with a cryogenically cooled, high-resolution x-ray spectrometer based on a 141 μm×141 μm Nb-Al-Al2O3-Al-Nb superconducting tunnel junction (STJ) detector in a demonstration experiment. Using monochromatized synchrotron radiation we studied the energy resolution of this energy-dispersive spectrometer for soft x rays with energies between 70 and 700 eV and investigated its performance at count rates up to nearly 60 000 cps. At count rates of several 100 cps we achieved an energy resolution of 5.9 eV (FWHM) and an electronic noise of 4.5 eV for 277 eV x rays (the energy corresponding to C K). Increasing the count rate, the resolution 277 eV remained below 10 eV for count rates up to ∼10 000 cps and then degraded to 13 eV at 23 000 cps and 20 eV at 50 000 cps. These results were achieved using a commercially available spectroscopy amplifier with a baseline restorer. No pile-up rejection was applied in these measurements. Our results show that STJ detectors can operate ...