Influence of fluoride ion on the electrochemical behaviour of β-Ti alloy for dental implant application
Influence of fluoride ion on the electrochemical behaviour of β-Ti alloy for dental implant application
复制标题
DOI:
10.1016/j.corsci.2010.01.008
复制
发表时间:
2010-05-01
影响因子:
8.3
通讯作者:
Kumar, S. Saravana
中科院分区:
文献类型:
--
作者:
Kumar, Satendra;Narayanan, T. S. N. Sankara;Kumar, S. Saravana
The corrosion behaviour of Ti-15Mo alloy in 0.15 M NaCl containing 0.01, 0.03, 0.06 and 0.5 M NaF is evaluated and its protective ability is compared with that of CP-Ti and Ti-6Al-4V alloy, to ascertain their suitability for dental implant application. The steady state current density of CP-Ti and, Ti-15Mo and Ti6Al-4V alloys in 0.15 M NaCl containing 0.03 M NaF at 200 mV vs. SCE is found to be 1, 2 and 6 mu A/cm(2), respectively, which indicate that all of them could offer a better corrosion resistance in the potential range that could exists in the oral environment. (C) 2010 Elsevier Ltd. All rights reserved.