Radiation dose from cone beam computed tomography for image-guided radiation therapy
Radiation dose from cone beam computed tomography for image-guided radiation therapy
复制标题
DOI:
10.1016/j.ijrobp.2007.08.062
复制
发表时间:
2008-01-01
影响因子:
7
通讯作者:
Lam, Nelson
中科院分区:
文献类型:
--
作者:
Kan, Monica W. K.;Leung, Lucullus H. T.;Lam, Nelson
Purpose: To perform a comprehensive study on organ absorbed doses and effective doses from cone beam computed tomography (CBCT) for three different treatment sites.Methods and Materials: An extensive set of dosimetric measurements were performed using a widely used CBCT system, the On-Board Imager (OBI). Measurements were performed using a female anthropomorphic phantom with thermoluminescent dosimeters (TLD). The effective doses to the body and the absorbed doses to 26 organs were reported using two different technical settings, namely, the standard mode and the low-dose mode. The measurements were repeated for three different scan sites: head and neck, chest, and pelvis. Comparisons of patient doses as well as image quality were performed among the standard mode CBCT, low-dose mode CBCT, and fan beam CT.Results: The mean skin doses from standard mode CBCT to head and neck, chest and pelvis were 6.7, 6.4, and 5.4 cGy per scan, respectively. The effective doses to the body from standard mode CBCT for imaging of head and neck, chest, and pelvis were 10.3, 23.7, and 22.7 mSv per scan, respectively. Patient doses from low-dose mode CBCT were approximately one fifth of those from standard mode CBCT.Conclusions: Patient position verification by standard mode CBCTacquired by OBI on a daily basis could increase the secondary cancer risk by up to 2% to 4%. Therefore lower mAs settings for daily CBCT should be considered, especially when bony anatomy is the main interest. (C) 2008 Elsevier Inc.