RF inhomogeneity compensation in structural brain imaging
RF inhomogeneity compensation in structural brain imaging
复制标题
DOI:
10.1002/mrm.10050
复制
发表时间:
2002-02-01
影响因子:
3.3
通讯作者:
Turner, R
中科院分区:
文献类型:
--
作者:
Deichmann, R;Good, CD;Turner, R
Three-dimensional T-1-weighted magnetization-prepared rapid gradient-echo (MP-RAGE) sequences with centric phase encoding (PE) in the inner loop provide structural brain images with a high spatial resolution and high tissue contrast. A disadvantage of this sequence type is the susceptibility to inhomogeneities of the radiofrequency (RF) coil, which may result in poor image contrast in some peripheral regions. A special excitation pulse is presented which compensates for these effects in both the head/foot and anterior/posterior directions. This pulse has a duration of only 1.3 ms and is thus compatible with the short repetition times (TRs) required for MP-RAGE imaging. It is shown experimentally that images acquired with the compensation pulse may be segmented without using intensity correction algorithms during data postprocessing. Magn Reson Med 47:398-402, 2002. (C) 2002 Wiley-Liss, Inc.