Design and Testing of an Advanced Implantable Neuroprosthesis With Myoelectric Control

Design and Testing of an Advanced Implantable Neuroprosthesis With Myoelectric Control
复制标题

DOI:
10.1109/tnsre.2010.2079952
复制
发表时间:
2011-02-01
影响因子:
4.9
通讯作者:
Peckham, P. Hunter
Peckham, P. Hunter
中科院分区:
工程技术2区
文献类型:
--
作者:
Hart, Ronald L.;Bhadra, Niloy;Peckham, P. Hunter

文献摘要

被引文献

相似文献

An implantable stimulator-telemeter (IST-12) was developed for applications in neuroprosthetic restoration of limb function in paralyzed individuals. The IST-12 provides 12 stimulation channels and two myoelectric signal (MES) channels. The MES circuitry includes a two-channel multiplexer, preamplifier, variable gain amplifier/bandpass filter, full-wave rectifier, and bin integrator. Power and control signals are transmitted from an external control unit to the IST-12 through an inductive link. Recorded MES signals are telemetered back to the external control unit through the same inductive link. Following bench testing, one device was implanted chronically in a dog for 15 months and evaluated. Conditions were identified in which MES could be recorded with minimal stimulus artifact. The ability to record MES in the presence of stimulation was verified, confirming the potential of the IST-12 to be used as a myoelectric controlled neuroprosthesis.