THE UNRELIABILITY OF CT-NUMBERS AS ABSOLUTE VALUES
THE UNRELIABILITY OF CT-NUMBERS AS ABSOLUTE VALUES
复制标题
DOI:
10.2214/ajr.139.3.443
复制
发表时间:
1982-01-01
影响因子:
5
通讯作者:
HATTERY, RR
中科院分区:
文献类型:
--
作者:
LEVI, C;GRAY, JE;HATTERY, RR
The use of CT [computed tomography] numbers as absolute values was examined by scanning a standard phantom on 5 CT scanners under a variety of conditions simulating those encountered in routine body CT scanning. There are significant differences in absolute CT numbers between most scanners (only 1 scanner produced CT numbers that were equal to zero for water). There are significant differences in absolute CT numbers between two scanners of the same manufacturer and model that were examined. There is a significant difference in CT numbers in a single phantom scan, dependent on location in the scan, and the format of this variability is not constant from one scanner to another. There may be a significant difference in absolute CT numbers, depending on various physical factors (e.g., kilovoltage, phantom orientation in scan aperture and position of the phantom in the scan aperture). The findings suggest that there is a wide range of CT numbers observed for a given tissue type as a result of scanner performance alone, and that if absolute CT numbers are to be used for diagnosis the user must document that these machine-related variations are less than the differences thought to be significant.