A compact neural recording interface based on silicon microelectrode
A compact neural recording interface based on silicon microelectrode
复制标题
DOI:
10.1007/s11431-013-5359-2
复制
发表时间:
2013-09
期刊:
影响因子:
--
通讯作者:
Jianqiang Han;Xu Zhang;Weihua Pei;Q. Gui;Ming Liu;Hongda Chen
中科院分区:
文献类型:
--
作者:
Jianqiang Han;Xu Zhang;Weihua Pei;Q. Gui;Ming Liu;Hongda Chen
A prototype of hybrid neural recording interface has been developed for extracellular neural recording. It consists of a silicon-based plane microelectrode array and a CMOS low noise neural amplifier chip. The neural amplifier chip is designed and implemented in 0.18 μm N-well CMOS 1P6M technology. The area of the neural preamplifier is only 0.042 mm2with a gain of 48.3 dB. The input equivalent noise is 4.73 μVrms within pass bands of 4 kHz. To avoid cable tethering for high dense multichannel neural recording interface and make it compact, flip-chip bonding is used to integrate the preamplifier chip and the microelectrode together. The hybrid device measures 3 mm×5.5 mm×330 μm, which is convenient for implant orin-vivoneural recording. The hybrid device was testified inin-vivoexperiment. Neural signals were recorded from hippocampus region of anesthetized Sprague Dawley rats successfully.