Comparison on radiation effective dose and image quality of right coronary artery on prospective ECG-gated method between 320 row CT and 2nd generation (128-slice) dual source CT

Comparison on radiation effective dose and image quality of right coronary artery on prospective ECG-gated method between 320 row CT and 2nd generation (128-slice) dual source CT
复制标题

DOI:
10.1002/acm2.12911
复制
发表时间:
2020-06-08
影响因子:
2.1
通讯作者:
Zhang, Jian-Hua
Zhang, Jian-Hua
中科院分区:
医学4区
文献类型:
--
作者:
Li, Ren-Feng;Hou, Chang-Long;Zhang, Jian-Hua

文献摘要

被引文献

相似文献

本回顾性研究旨在比较320排计算机断层扫描(CT)和第二代(128层)双源CT在前瞻性ECG门控方法上的右冠状动脉(RCA)图像质量和有效辐射剂量。共215名候选人采用前瞻性心电门控方法进行了CT冠状动脉造影,320排CT组120例,双源CT组95例。我们将RCA图像质量评分分为1/2/3/4,这意味着极好/良好/足够/不可评估,并考虑了心率以及辐射剂量。320排CT与第二代双源CT的RCA图像质量评分1/2无统计学差异,但心率较低(
This retrospective study was to compare the image quality of right coronary artery (RCA) and effective radiation dose on prospective ECG-gated method between 320 row computed tomography (CT) and 2nd generation (128-slice) dual source CT. A total of 215 candidates underwent CT coronary angiography using prospective ECG-gated method, 120 patients enrolled in 320 row CT group, and 95 patients in dual source CT group. We divided RCA image quality scores as 1/2/3/4, which means excellent/good/adequate/not assessable and heart rates were considered, as well as the radiation dose. There is no statistically significant difference of RCA image quality of Score 1/2 between 320 row CT and 2nd generation dual source CT, but lower heart rate (