APPLICATION OF DYNAMIC CAPILLARY ISOELECTRIC-FOCUSING TO THE ANALYSIS OF HUMAN HEMOGLOBIN-VARIANTS
APPLICATION OF DYNAMIC CAPILLARY ISOELECTRIC-FOCUSING TO THE ANALYSIS OF HUMAN HEMOGLOBIN-VARIANTS
复制标题
DOI:
10.1002/elps.1150150105
复制
发表时间:
1994-01-01
期刊:
影响因子:
2.9
通讯作者:
THORMANN, W
中科院分区:
文献类型:
--
作者:
MOLTENI, S;FRISCHKNECHT, H;THORMANN, W
Capillary isoelectric focusing (CIEF) with electroosmotic zone displacement of normal and pathological hemoglobins (Hb) is reported. CIEF is performed in untreated, open-tubular, fused-silica capillaries of 75 mu m internal diameter using methylcellulose for dynamic column conditioning. After direct injection of hemolysates mixed with carrier ampholytes, high resolution separation of Hb variants, including Hb A(1c), A, F,D, S, E and A(2), is obtained, this permitting unambiguous characterization of Hb patterns of normal adults, newborns, patients with diabetes, different hemoglobinopathies and thalassemia syndromes. Qualitatively, the CIEF data compare well with those obtained by gel isoelectric focusing and high-performance liquid chromatography. CIEF is demonstrated to be a simple, rapid and fully instrumental approach to Hb analysis. Run times of less than 20 min make CIEF an attractive method for routine Hb investigations and screening programs.