A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles
A novel technique for measuring the low-dose envelope of pencil-beam scanning spot profiles
复制标题
DOI:
10.1088/0031-9155/58/12/n171
复制
发表时间:
2013-06-21
影响因子:
3.5
通讯作者:
McDonough, James E.
中科院分区:
文献类型:
--
作者:
Lin, Liyong;Ainsley, Christopher G.;McDonough, James E.
To investigate the profile measurement capabilities of an IBA-Dosimetry scintillation detector and to assess its feasibility for determining the low-intensity tails of pencil-beam scanning spots, the responses of the scintillation detector and Gafchromic EBT2 film to a 115 MeV proton spot were measured in-air at the isocenter. Pairs of irradiations were made: one lower-level irradiation insufficient to cause saturation, and one higher-level irradiation which deliberately saturated the central region of the spot, but provided magnification of the tails. By employing the pair/magnification technique, agreement between the film and scintillation detector measurements of the spot profile can be extended from 4% of the central spot dose down to 0.01%. Gamma analysis between these measurements shows 95% and 99% agreement within a +/- 9 cm bound using criteria of 3 mm/3% and 5 mm/5%, respectively. Above 4%, our 115 MeV proton spot can be well-described by Gaussian function; below 4%, non-Gaussian, diamond-shaped tails predominate.