A field-sweep/field-lock system for superconducting magnets - Application to high-field EPR
A field-sweep/field-lock system for superconducting magnets - Application to high-field EPR
复制标题
DOI:
10.1016/j.jmr.2006.09.012
复制
发表时间:
2006-12-01
影响因子:
2.2
通讯作者:
Griffin, Robert G.
中科院分区:
文献类型:
--
作者:
Maly, Thorsten;Bryant, Jeff;Griffin, Robert G.
We describe a field-lock/field-sweep system for the use in superconducting magnets. The system is based on a commercially available field mapping unit and a custom designed broad-band H-1 NMR probe. The NMR signal of a small water sample is used in a feedback loop to set and control the magnetic field to high accuracy. The current instrumental configuration allows field sweeps of +/- 0.4 T and a resolution of up to 10(-5) T (0.1 G) and the performance of the system is demonstrated in a high-field electron paramagnetic resonance (EPR) application. The system should also be of utility in other experiments requiring precise and reproducible sweeps of the magnetic field such as DNP, ENDOR or PELDOR. (c) 2006 Elsevier Inc. All rights reserved.