ADMITTANCE MODELS FOR OPEN-ENDED COAXIAL PROBES AND THEIR PLACE IN DIELECTRIC-SPECTROSCOPY

ADMITTANCE MODELS FOR OPEN-ENDED COAXIAL PROBES AND THEIR PLACE IN DIELECTRIC-SPECTROSCOPY
复制标题

DOI:
10.1088/0031-9155/39/12/004
复制
发表时间:
1994-12-01
影响因子:
3.5
通讯作者:
GRANT, EH
GRANT, EH
中科院分区:
工程技术2区
文献类型:
--
作者:
GABRIEL, C;CHAN, TYA;GRANT, EH

文献摘要

被引文献

相似文献

Starting from a rigorous formulation for the admittance of an open ended coaxial probe, this paper identifies the simplifying assumptions that are introduced to obtain more practical solutions. The predictions of the models are compared to each other and to experimental data obtained on an independently calibrated experimental set-up. The advantages and limitations of a model are determined by the theoretical derivation and the numerical solution. The model identified as most suitable for the measurement of the dielectric properties of biological materials is used to determine the relative permittivity and conductivity of muscle and skull bone in the frequency range 1 MHz to 20 GHz.