B1-insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T
B1-insensitive T2 preparation for improved coronary magnetic resonance angiography at 3 T
复制标题
DOI:
10.1002/mrm.20835
复制
发表时间:
2006-04-01
影响因子:
3.3
通讯作者:
Pettigrew, RI
中科院分区:
文献类型:
--
作者:
Nezafat, R;Stuber, M;Pettigrew, RI
At 3 T, the effective wavelength of the RF field is comparable to the dimension of the human body, resulting in B-1 standing wave effects and extra variations in phase. This effect is accompanied by an increase in B-0 field inhomogeneity compared to 1.5 T. This combination results in nonuniform magnetization preparation by the composite MLEV weighted T-2 preparation (T-2 Prep) sequence used for coronary magnetic resonance angiography (MRA). A new adiabatic refocusing T-2 Prep sequence is presented in which the magnetization is tipped into the transverse plane with a hard RF pulse and refocused using a pair of adiabatic fast-passage RF pulses. The isochromats are subsequently returned to the longitudinal axis using a hard RF pulse. Numerical simulations predict an excellent suppression of artifacts originating from B, inhomogeneity while achieving good contrast enhancement between coronary arteries and surrounding tissue. This was confirmed by an in vivo study, in which coronary MR angiograms were obtained without a T-2 Prep, with an MLEV weighted T-2 Prep and the proposed adiabatic T-2 Prep. Improved quantitative and qualitative coronary MRA image measurement was achieved using the adiabatic T-2 Prep at 3 T.