Precision of extracting absorption profiles from weakly scattering media with spectroscopic time-domain optical coherence tomography
Precision of extracting absorption profiles from weakly scattering media with spectroscopic time-domain optical coherence tomography
复制标题
DOI:
10.1364/opex.12.001677
复制
发表时间:
2004-04-19
期刊:
影响因子:
3.8
通讯作者:
Drexler, W
中科院分区:
文献类型:
--
作者:
Hermann, B;Bizheva, K;Drexler, W
The feasibility of spectroscopic optical coherence tomography (SOCT) to quantify spatially localized absorption profiles of chromophores embedded in weakly scattering media with a single measurement over the full spectral bandwidth of the light source was investigated by using a state-of-the-art ultra-broad bandwidth Ti: Al2O3 laser (lambda(c) = 800 nm, Deltalambda = 260 nm, P-out = 120 mW ex- fiber). The precision of the method as a function of the chromophore absorption, the sample thickness, and different parameters related to the measurement procedure was evaluated both theoretically and experimentally in single and multilayered phantoms. It is demonstrated that in weakly scattering media SOCT is able to extract mu(a)(lambda) as small as 0.5 mm(-1) from 450 mum thick phantoms with a precision of similar to2% in the central and similar to8% at the edges of the used wavelength region. As expected, in phantoms with the same absorption properties and thickness similar to180 mum the precision of SOCT decreases to >10% in the central wavelength region. (C) 2004 Optical Society of America.