Clinical evaluation of a breathing protocol for PET/CT

Clinical evaluation of a breathing protocol for PET/CT
复制标题

DOI:
10.1007/s00330-003-2174-3
复制
发表时间:
2004-06-01
期刊:
影响因子:
5.9
通讯作者:
Goerres, GW
Goerres, GW
中科院分区:
医学2区
文献类型:
--
作者:
de Juan, R;Seifert, B;Goerres, GW

文献摘要

被引文献

相似文献

本研究的目的是评估在修改CT扫描的呼吸方案之前和之后,在联合PET/CT扫描仪上采集的共配准正电子发射断层扫描/计算机断层扫描(PET/CT)图像上呼吸诱导的曲线呼吸伪影(RICA)的频率和严重程度,回顾性分析了两组100例患者,在执行呼吸协议之前和之后,在正常呼气位置屏气(BH)以采集CT图像。CT数据被用作衰减图和图像配准。由两名观察员一致使用0 - 3的量表对共配准PET/CT和PET图像(包括最大强度投影)进行评级。0表示RICA不可见,1、2和3描述了严重程度增加的伪影。两组RICA发生率有显著性差异(p
The aim of this study was to assess the frequency and severity of respiration-induced curvilinear respiration artifacts (RICA) on co-registered positron emission tomography/computed tomography (PET/CT) images acquired on a combined PET/CT scanner before and after modifying the respiration protocol for CT scanning, with retrospective analysis of two groups of 100 patients each, before and after implementing a respiration protocol with breath-hold (BH) in the normal expiration position for the acquisition of the CT images. The CT data were used as attenuation map and for image co-registration. A ranking of co-registered PET/CT and PET images (including maximum intensity projection) was done by two observers in consensus using a scale from 0 to 3. Zero indicated that no RICA was visible and 1, 2, and 3 described artifact with increasing severity. A significant difference in RICA occurrence was found between the two groups (p